Calibration of heat:
What type of standard do you need for the calibration of heat and temperature of DSC?
Answer:
It is recommended to use pure metal standards (99.999% pure) for the temperature and heat calibration of DSC and DTA. According to the manufacturers of thermal analysis equipment, one or more standards are needed. Indium (In) is the commonly used standard for the calibration of temperature and peak area.