Determination of ascorbic acid (Vitamin C) within the citrus juice through the standard addition and calibration curve methods:
Principle: Ascorbic acid gives a well defined polarographic oxidation wave.
Use freshly prepared diluted juice
Calibration curve method
Prepare a fresh stock solution of 50 cm3 of 0.2% ascorbic acid. Arrange 5 standard solutions of ascorbic acid within volumetric flasks of 25 cm3. To everyone add 0.5 cm3 of 0.5 M acetate buffer and different volumes of 0.2% ascorbic acid, 0, 200, 400, 600 and 800 µdm3. Dilute to the mark along with distilled water. For every solution record polarograms over the potential range -150 to + 200 mV vs Ag/AgCl/1M KCl reference electrode.
Plot id vs c of ascorbic acid (calibration curve).