Pipettor calibration check

Assignment Help Chemistry
Reference no: EM131128955

Experiment 1a: Pipettor Calibration Check

Question:Are my electronic pipettors operational?

1. Select any 3 of the 5 electronic pipettors within your workstation.
2. Select 5 volume settings within the specified range on the pipettor. Measure those volumes of deionized water in a weigh boat and record the corresponding weight in your laboratory notebook. (Hint: A table would be appropriate for data collection.)
3. Use the density of water to correlate what the theoretical weight measurement should be.
4. Using Excel, graph the weight of water measured (y-axis) as a function of the volume setting on the pipettor (x-axis). Label the axes, include your R2 value and secure the data in your lab notebook BEFORE leaving class.
5. Note in your notebook whether your pipettors are operational or not.

Experiment 1b. Bradford Protein Concentration Assay

Questions:
• What is the concentration of the protein sample provided?
• What are the limits in which the protein concentration could be determined experimentally?
• How linear is the standard curve?
• How accurate is the standard curve?

The Beer-Lambert Law (A = εlc) can be useful in estimating the total concentration of a single protein in solution. However, the total protein concentration in the Beer-Lambert law is based on the absorbance of a single amino acid, tryptophan, at 280 nm. What if only some proteins in the mixture contain Trp?

Then, the measured absorbance would not be proportional to the actual concentration of proteins in the mixture. Instead, a better estimate of total protein concentration can be determined from the Bradford Protein Assay.

The Bradford Protein Assay is one of several colorimetric assays commonly used to determine total protein concentration in a pure sample or in a mixture. The method is based on the proportional binding of the dye Coomassie (present in the Bradford Protein Assay Reagent) to proteins. Within the linear range of the assay, the more protein present, the more Coomassie binds. Furthermore, as the protein concentration increases, the color of the test sample becomes more intensely blue.

Many absorption assays are endpoint assays - meaning some reactant is added lastly to each sample to stop the reaction. The Bradford Protein Assay is instead a continuous assay; the longer the incubation, the greater the net binding of Coomassie dye to any protein present in the sample and the higher the absorbance readings. Since the assay does not measure an absolute endpoint, the concentration of protein in the test sample must be deduced by comparison with a set of standard proteins of known concentration. The assumption is that the protein in the test sample will give an identical response to that of the standard protein at the same concentration.

Coomassie dye in the Bradford Protein Assay absorbs light at 595 nm. The protein concentration of a test sample is determined by comparison to that of a series of protein standards known to reproducibly exhibit a linear absorbance profile in this assay. Although different protein standards can be used, we have chosen the most widely used protein as our standard - Bovine Serum Albumin (BSA).

1. Prepare 200 µL of a 4-fold dilution of 2 mg/mL BSA.

Add ______ µL of 2 mg/mL BSA + ______  µL of dI water = 200 µL of ______ mg/mL BSA (working solution).

2. Mix thoroughly but gently. The working solution of BSA will be used for the remainder of this experiment. Note: If you are using BSA with a different initial concentration than 2 mg/mL, adjust the dilution as needed.

3. Generate samples according to Table 1 below in disposable cuvettes. Initially, mix only the protein and water in each cuvette.

4. Noting the time, add the Bradford Protein Assay reagent to each cuvette at the same pace with which you would read the absorbance measurements. For example, you should add Bradford Protein Assay Reagent to each cuvette in 30-sec intervals because you can measure and record absorbance data comfortably within 30-sec intervals. Use longer intervals, if necessary.

5. Note that a dilution of the "unknown protein" sample may be required for the resulting absorbance to fall within the linear range of the assay. Ask your TA or instructor for guidance on when to dilute your sample. In real experimental settings, you might have to determine the need for dilutions by trial and error.

Table 1. Bradford Protein Assay Setup Using a Stock BSA Solution at 0.5 mg/mL

1281_1.png

6. Allow each sample to incubate at room temperature for a few (i.e. 2-10) minutes. (Record the actual incubation time in your notebook. Never allow the reaction to proceed for so long that the absorbance measurements grossly exceed 1.0.) Meanwhile, turn on your laptop and connect the SpectroVis Plus instrument to it via the USB connector.

7. Launch the LoggerPro software.

8. Select Experiment à Change Units à Spectrometer 1 à Absorbance

9. To calibrate the instrument:

a. Choose, Experiment à Calibrate à Spectrometer
b. Fill a cuvette about ¾ full with dI water
c. Place the water blank into the spectrometer
d. Follow the instructions in the dialog box to complete the calibration, then click OK.

10. Select the icon for Configure Spectrometer 1 Data Collection

a. Collection mode: Absorbance vs. concentration
b. Column Name: Concentration
c. Units: µg/mL
d. Single 10 nm band
e. Click 594.8 nm (for an estimate of 595 nm)
f. Click OK

11. Insert the sample into the SpectroVis Plus instrument.

12. For data collection:

a. Click Collect
b. Click Keep
c. Enter the concentration of the sample
d. Click Stop
e. Remove the sample
f. Insert another sample
g. At the appropriate time interval, click Collect
h. Select Append to Latest Data
i. Repeat until all data have been collected.

13. To save data, click File à Export As à CSV. Enter a file name and save the data in an appropriate directory.

14. Open the CSV file in EXCEL. To process the data:

a. Subtract the absorbance of the REFERENCE from each BSA standard curve data point.
b. Generate a scatter plot of the Absorbance at 595 nm (on the y-axis) as a function of the [BSA], µg/mL (on the x-axis).
c. Label the axes.
d. Remove excess horizontal lines
e. Determine the linear range of the data. Use only the linear range (use at least 5 points preferably) to deduce the equation of the linear fit.
f. Add the linear trendline fit, showing the equation and R2 value on the graph
g. Print the data and tape or glue it into your laboratory notebook.
h. Use the linear equation to compute the concentration of the protein unknown.
i. To assess the accuracy of the assay, put the absorbance data for each BSA standard back into the linear equation to compute the BSA concentration for each standard. Compare the computed BSA concentration to the theoretical concentration.

Experiment 1c. DNA Sample Concentration Determination

Question: What is the concentration of the unknown DNA sample?

1. Add 500 µL of the purified DNA solution to 500 µL of 50 mM Tris-HCl at pH 8 directly into a Plastibrand® cuvette, which is transparent to low range UV light.

2. Measure the absorbance of the DNA solution at 260 nm by following the steps below and record the absorbance in your lab notebook.
• Turn on the Shimadzu UV-1601 spectrometer
• Wait for all system checks to complete (usually in 5 min). If necessary, turn the instrument off and back on.
• Press MODE to see all measurement options.
• Select the PHOTOMETRIC mode (1).
• Make sure that data is shown in absorbance units rather than %T. If necessary press the appropriate function key (F1, F2, F3 or F4) to change the unit of measurement.
• Insert the reference and blank cuvettes into the instrument and close the sample chamber door.
• Press AUTOZERO and wait for a beep. The absorbance should be 0.0000.
• Remove the cuvette from the Sample (front) position.
• Cover the cuvette containing the DNA sample and mix with inversion.
• Insert the DNA sample into the instrument and close the sample chamber door.
• Record the absorbance reading.
• Remove all cuvettes and turn the instrument off.

3. As described by Sambrook and coworkers in the reference below, determine the concentration (in µg/mL) of the purified DNA sample as follows:

[DNA]original (µg/mL) = (A260) ? (50 µg/mL) ? (Dilution Factor)

Attachment:- bradford.xlsx

Reference no: EM131128955

Questions Cloud

The poisson distribution with the parameter : Show that if X is a random variable having the Poisson distribution with the parameter λ and λ→∞, then the moment-generating function of that is, that of a standardized Poisson random variable, approaches the moment-generating function of the sta..
Complete swot analysis and swot matrix : Provide a Situational Analysis of the current Internal and Customer Environments, Research and prepare an analysis of the External Environment and Complete SWOT Analysis and SWOT Matrix.
What is the difference between a market order and a limit : What is the difference between a market order and a limit order? What is the potential downside to each type of order?
What is the difference between a hostile takeover and merger : What is the difference between a hostile takeover and a merger?
Pipettor calibration check : Are my electronic pipettors operational? -  What are the limits in which the protein concentration could be determined experimentally?
Random variable having the uniform density : A point D is chosen on the line AB, whose midpoint is C and whose length is a. If X, the distance from D to A, is a random variable having the uniform density with α = 0 and β = a, what is the probability that AD, BD, and AC will form a triangle?
Why would an investment banker advise a firm to issue : Why would an investment banker advise a firm to issue a security using best efforts rather than underwriting?
The daily consumption of electric power : In a certain city, the daily consumption of electric power in millions of kilowatt-hours can be treated as a random variable having a gamma distribution with α = 3 and β = 2.
Describe how marketing efforts and marketing mix will change : In 530 words please describe how marketing efforts and marketing mix will change with each phase in the product life cycle of a fictitious summer makeup kit from MAC Cosmetics' that focuses that includes a bronzer, highlighter, blush, BB Cream, li..

Reviews

Write a Review

Chemistry Questions & Answers

  Explain how much work is needed to send 2.5 gal/hr of fuel

How much work is needed to send 2.5 gal/hr of fuel through a 8mm line with a .09mm nozzle at 1500 psi absolute coming from a fuel tank at 1 atm absolute

  How to calculate the equilibrium concentration of h2

Suppose at some temperature, Kc= 6.20 X 10^8 for the reaction 2NO(g) + 2H2(g) N2 (g) + 2H2O(g). 5.00 M NO(g) and 5.00 M H2 (g) are placed in a reaction flask. Calculate the equilibrium concentration of H2 (g)

  Explain a solution of h2co3 and hco3 will change the ph

A buffering system in the blood is H2CO3/HCO4. At the normal blood pH (7.4) a solution of H2CO3 and HCO3 will change the pH abruptly if small amounts of acid or base are added since the pKa of H2CO3 is 6.4.

  Dilute solution to a concentration of 5 g/ml

Problem- If I have 300mL of a solution of 25 g/mL. I want to dilute it to a concentration of 5 g/mL. What do I do? Please illustrate work and any conversions to assist me better understand.

  Use a sketch to show the change that occurs as a sulfur

use a sketch to show the change that occurs as a sulfur atombecomes a sulfide ion. describe how the distance from the

  Describe the molar mass of oxygen

A particular metallic element, M, has a heat capacity of 0.36 J * g^-1 * K^-1. It forms an oxide that contains 2.90 g of M per gram of oxygen. (a) Using the law of Dulong and Petit (Cp x M = 25 J per mole per Kelvin), estimate the molar mass of th..

  What quantity of work associated with the compression of gas

Calculate the quantity of work associated with the compression of gas from 5.60 L to 3.40 L at a constant pressure of 1.23 atm.

  A sample of n2 gas occupies 240 l at 20c if the gas is in

1. a sample of n2 gas occupies 2.40 l at 20?c. if the gas is in acontainer that can contract or expand at constant

  Explain is the reaction mixture in equilibrium

Question- A reaction mixture contains 0.114 atm I2, 0.102 atm Cl2 and 0.355 atm ICl. Is the reaction mixture in equilibrium. If not, which direction will the reaction proceed

  Considering the given reaction 2s3o2rarr2so3 in which

considering the given reaction 2s3o2rarr2so3 in which elemental s reacts with oxygen to produce so3 then find out how

  Compute the kcat in sec-1 for the enzyme

For the kinetic assay, 0.1ml of a .25mg/mL solution of enzyme was utilized and the enzyme has a molecular weight of 95,000 g/mole. Presume a reaction volume of 3mL. Compute the kcat (in sec-1) for the enzyme

  Explain what is the spin of the other electron

If the spin of one electron in an orbital is counterclockwise, what is the spin of the other electron in that orbital. Both clockwise and counterclockwise

Free Assignment Quote

Assured A++ Grade

Get guaranteed satisfaction & time on delivery in every assignment order you paid with us! We ensure premium quality solution document along with free turntin report!

All rights reserved! Copyrights ©2019-2020 ExpertsMind IT Educational Pvt Ltd